ARTIFICIAL CELLS BLOOD SUBSTITUTES AND BIOTECHNOLOGY, cilt.40, sa.4, ss.280-284, 2012 (SCI-Expanded)
In this paper, a novel amperometric choline biosensor with immobilization of choline oxidase on electrochemically polymerized polypyrrole-polyvinylsulphonate (PPy-PVS) film has been accomplished via the entrapment technique. The effects of pH and temperature were investigated and optimum parameters were found to be 9.0 and 60 degrees C, respectively. There are two linear parts in the region between 1.0 x 10(-7) -1.0 x 10(-6) M (R-2 = 0.997) and 1.0 x 10(-5) -1.0 x 10(-3) M (R-2 = 0.986). The storage stability and operation stability of the enzyme electrode were also studied.