True coincidence-summing corrections for the coincident gamma-rays measured with coplanar grid CdZnTe detectors


Yucel H., Solmaz A. N., Kose E., Bor D.

APPLIED RADIATION AND ISOTOPES, cilt.68, sa.6, ss.1040-1048, 2010 (SCI-Expanded) identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 68 Sayı: 6
  • Basım Tarihi: 2010
  • Doi Numarası: 10.1016/j.apradiso.2010.01.042
  • Dergi Adı: APPLIED RADIATION AND ISOTOPES
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.1040-1048
  • Gazi Üniversitesi Adresli: Hayır

Özet

In this study, true coincidence-summing (TCS) correction factors have been measured for the sources (22)Na, (60)Co, (133)Ba and (152)Eu by use of three large volume coplanar grid CdZnTe (acronym: CZT) detectors. In case of a close-in detection geometry, two different TCS calculation algorithms were used to compute the required TCS correction factors. Both of the algorithms are based on the measured total-to-peak (TTP) ratio and full-energy peak (FEP) efficiency values that were obtained using almost "single" energy and coincidence-free nuclides. The results for TCS correction factors obtained by two different algorithms were agreeable to each other. The obtained TCS factors were ranged from about 7% to 30.5% in a 2250 mm(3) CZT detector when a close counting geometry was used. For other two detectors with a volume of 1000 and 1687.5 mm(3), the resulted TCS correction factors were relatively smaller and varied between about 0.1% and 20% at the close counting geometry condition. Therefore, the results indicate that there is a need for the estimation of TCS corrections in CZT detectors, especially when their crystal volumes are greater than 1 cm(3) and these detectors are used in the case of a close-in detection geometry. (C) 2010 Elsevier Ltd. All rights reserved.