PHYSICS IN MEDICINE AND BIOLOGY, vol.59, no.22, pp.6749-6758, 2014 (SCI-Expanded)
The energy dependence of the response of BC-408 plastic scintillator (PS), an approximately water-equivalent material, has been investigated by employing standardized x-ray beams. IEC RQA and ISO N series x-ray beam qualities, in the range of 40-100 kV(p), were calibrated using a PTW-type ionization chamber. The energy response of a thick BC-408 PS detector was measured using the multichannel pulse height analysis method.