Rehabilitation of the Severely Atrophic Maxilla with Subperiosteal Implants: A Biomechanical and Decision Analysis of Material and Configuration Choices
Biomimetics, cilt.11, sa.6, 2026 (SCI-Expanded, Scopus)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 11 Sayı: 6
- Basım Tarihi: 2026
- Doi Numarası: 10.3390/biomimetics11060433
- Dergi Adı: Biomimetics
- Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Compendex, Directory of Open Access Journals, Academic Search Ultimate (EBSCO), Natural Science Collection (ProQuest), Biological Science Database (ProQuest)
- Anahtar Kelimeler: CFR-PEEK, finite element analysis, subperiosteal implant, titanium, TOPSIS
- Gazi Üniversitesi Adresli: Evet
Özet
Background/Objectives: Patient-specific subperiosteal implants are increasingly used to treat severely atrophic ridges due to advances in digital planning and additive manufacturing. This study aimed to evaluate the effects of material type and implant configuration on stress distribution in subperiosteal implant systems and to compare their overall biomechanical performance using a multi-criteria decision framework. Methods: A three-dimensional model of a severely atrophic maxilla was reconstructed to simulate four clinical scenarios combining two configurations (one-piece and two-piece) and two materials (titanium and 60% carbon fiber-reinforced polyetheretherketone). Finite element analysis was conducted to assess stress distribution within the implant body, fixation screws, prosthetic framework, and surrounding bone under vertical and oblique loading conditions. Maximum and minimum principal stresses were evaluated in bone, whereas von Mises stresses were calculated for implant components. The resulting biomechanical indicators were subsequently integrated using an entropy weight–TOPSIS multi-criteria decision analysis. Results: Principal stresses in the surrounding bone showed minimal variation between titanium and 60% carbon fiber-reinforced polyetheretherketone across all configurations. Implant configuration had a more pronounced effect on implant body stress. Under oblique loading, the two-piece configuration demonstrated substantially higher implant stresses than the one-piece design, whereas under vertical loading, lower implant stresses were observed in the two-piece configuration. The multi-criteria analysis ranked the one-piece titanium model highest under oblique loading and the two-piece titanium model highest under vertical loading. Conclusions: Implant configuration and loading direction influenced biomechanical behavior more than material selection in patient-specific subperiosteal implants.