Determination of fumonisins B-1 and b2 in corn by liquid chromatography/mass spectrometry with immunoaffinity column cleanup: Single-laboratory method validation


Senyuva H. Z., ÖZCAN KABASAKAL S., Cimen D., Gilbert J.

JOURNAL OF AOAC INTERNATIONAL, cilt.91, sa.3, ss.598-606, 2008 (SCI-Expanded) identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 91 Sayı: 3
  • Basım Tarihi: 2008
  • Dergi Adı: JOURNAL OF AOAC INTERNATIONAL
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.598-606
  • Gazi Üniversitesi Adresli: Evet

Özet

A single-laboratory method validation was conducted to establish the effectiveness of an immunoaffinity column cleanup procedure followed by liquid chromatography/mass spectrometry (LC/MS) for the determination of fumonisins B-1 and B-2 (FB1 + FB2) in corn. The test portion is extracted with acetonitrile-methanol-water (25 + 25 + 50). The extract is filtered, diluted with phosphate-buffered saline solution, and applied to an immunoaffinity column. FB1 + FB2 are removed with methanol and directly determined by reversed-phase LC with MS detection using selected-ion monitoring of 2 characteristic ions in each case. Test portions of blank corn samples were spiked with a mixture of FB1 + FB2 to give total levels of 200 and 500 ng/g, respectively. Recoveries of both FB1 and FB2 from spiked samples averaged 90.4-101%. Based on results for spiked raw corn (triplicates at 2 levels), the relative standard deviation for repeatability ranged from 2.8 to 7.1%. The accuracy of the method was demonstrated by analysis of Food Analysis Performance Assessment Scheme (FAPAS (R)) test material. The method was also applied to a small survey of processed corn products such as corn chips, cornflakes, and popcorn.