A new, fast and highly sensitive lead-sensitive solid-state electrode containing PbO2, Ag powder and CuS salts was prepared. The electrode was measured to be sensitive to 1.0 × 10–6 – 0.1 mol/dm3 Pb(II) concentration with a slope of 32 ± 0.5 mV. With this value, it showed a behaviour in accordance with the Nernst equation. The response time of the electrode was determined to be 10 s and the working range between pH 1 and 9. The interference effect of the electrode was investigated. Validation values of the electrode were measured. The LOD value of the electrode was calculated as 1.5 × 10−8 mol/dm3. This feature made the electrode important for the analysis of low concentration Pb(II). Determination of lead in drinking water was performed with both the electrode and ICP-OES and the results were compared.